Test Code LAB02936701 Blood Gases, Venous Cord Blood
Methodology
Electrode Methodology
Includes pH, pCO2, pO2, bicarbonate (HCO3), base excess, and % O2 saturation.
Performing Laboratory
CentraCare Laboratory Services
Specimen Requirements
Specimen Type: Venous cord blood
Container/Tube: Heparinized syringe
Note: Samples cannot be collected in microtainers. No capillary blood samples.
Collection Instructions:
1. Avoid exposure of specimen to atmosphere.
2. Hand deliver sample to CCLS lab. Do not send in tube station.
Specimen Transport Temperature
Ambient
Reference Values
pH: 7.30-7.40 units
Critical values (automatic call-back): ≤7.25 units or ≥7.60 units
pCO2: 33.9-48.3 mm Hg
Critical values (automatic call-back): ≤20 mm Hg or ≥70 mm Hg
pO2: 22.8-34.6 mm Hg
Critical value (automatic call-back): <30 or > 40 mm Hg
HCO3: 23-28 mmol/L
Critical values (automatic call-back): ≤14 mmol/L or ≥40 mmol/L
Base excess: -2.0 to 3.0 mmol/L
% O2 saturation: 60-80%
Day(s) Test Set Up
Monday through Sunday
Test Classification and CPT Coding
82803