Test Code LAB0002343 Phospho-Tau 217, Plasma
Additional Codes
Mayo Test ID |
---|
PT217 |
Ordering Guidance
This assay is useful for individuals presenting with mild cognitive impairment or early dementia. Results must be interpreted in conjunction with other diagnostic tools such as neurological examination, neurobehavioral tests, imaging, and routine laboratory tests.
This assay should not be ordered for individuals younger than 50 years or in cognitively unimpaired individuals regardless of age.
This assay should not be used to predict the development of dementia or other neurologic conditions.
Specimen Required
Collection Container/Tube: Lavender top (EDTA)
Submission Container/Tube: Plastic screw-top vial
Specimen Volume: 0.6 mL
Collection Information: Centrifuge and aliquot plasma into plastic vial. Do not submit in original tube.
Useful For
Evaluation of individuals, aged 50 years and older, presenting with cognitive impairment who are being assessed for Alzheimer disease and other causes of cognitive decline
This test is not intended as a screening test for Alzheimer disease in asymptomatic individuals.
Method Name
Chemiluminescent Enzyme Immunoassay
Reporting Name
Phospho-Tau(217), PSpecimen Type
EDTA PlasmaSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
EDTA Plasma | Refrigerated (preferred) | 14 days | |
Frozen | 90 days | ||
Ambient | 72 hours |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
Negative: ≤0.185 pg/mL
Intermediate: 0.186-0.324 pg/mL
Positive: ≥0.325 pg/mL
Interpretation
Negative: A normal (negative) phosphorylated Tau217 (p-Tau217) result is consistent with a negative (normal) amyloid-positron emission tomography (PET) scan result. This result indicates a reduced likelihood that an individual has neuropathological changes associated with Alzheimer disease.
Intermediate: An intermediate p-Tau217 result cannot accurately differentiate between the presence or absence of neuropathological changes associated with Alzheimer disease. Further testing, such as amyloid-positron emission tomography (PET) or cerebrospinal fluid Abeta42 and tau biomarkers, is needed to determine the likelihood of neuropathological changes associated with Alzheimer disease being present.
Positive: An elevated (positive) p-Tau217 result is consistent with a positive (abnormal) amyloid-positron emission tomography (PET) scan result. This result is consistent with the presence of neuropathological changes associated with Alzheimer disease. In the proper clinical context this test is supportive of Alzheimer disease being related to current clinical symptoms. This test has not been demonstrated to provide information on the risk of an asymptomatic individual developing symptoms related to Alzheimer disease in the future.
Clinical performance of this test was established in a study of 427 individuals aged 50 years and older with mild cognitive impairment or early dementia with a 64% prevalence of amyloid pathology defined by an amyloid-PET and a Centiloid scale value of 25 or more. For detection of an abnormal amyloid-PET, pTau217 test sensitivity at the lower cutpoint (≤0.185 pg/mL) was 92% and the specificity at the upper cutpoint (≥0.325 pg/mL) was 96%. The diagnostic performance of this test has not been established in asymptomatic individuals.
Day(s) Performed
Monday through Friday
Report Available
1 to 3 daysSpecimen Retention Time
3 monthsPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
83520
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
PT217 | Phospho-Tau(217), P | 104663-0 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
217PT | pTau217, P | 104663-0 |
PTINT | pTau217 Interpretation | 69048-7 |
Forms
If not ordering electronically, complete, print, and send a Neurology Specialty Testing Client Test Request (T732) with the specimen.